A long-term study of CR-39 detectors for neutron dosimetry

L. Tommasino, G. Torri, M. Cavaioli, M. Riccardi, Trinh Van Giap

Research output: Contribution to journalArticle

2 Citations (Scopus)

Abstract

In the past decade, the neutron-response of CR-39 dosimeters has been investigated several times using different etching processes and materials from different manufacturers. In spite of the fact that the data have been gathered using different batches, stored and aged under different conditions, the neutron-response has resulted highly consistent. By contrast, the detector background has shown lack of reproducibility not only from batch to batch but also from sheet to sheet within the same batch. © 1994.
Original languageEnglish
Pages (from-to)719 - 722
Number of pages4
JournalNuclear Tracks And Radiation Measurements (1993)
Volume22
Issue number1-4
DOIs
Publication statusPublished - 1993
Externally publishedYes