Effect of the encapsulating material on the peak3/peak5 response ratio of TLD-300 irradiated with neutrons of various energy

M. Angelone, P. Batistoni, M. Pillon

Research output: Contribution to journalArticle

1 Citation (Scopus)

Abstract

The analysis of experimental data available in the literature shows that the neutron sensitivity ratio m = k3/k5for the two main glow peaks P3and P5of TLD-300 (for ribbons of 3.2 × 3.2 × 0.9 mm3) is dependent upon the response ratio R3/R5of these two peaks. This finding is in agreement with predictions from an equation derived from the two-peak method equations. The result implies that the peak response ratio R3/R5for TLD-300 surrounded by Tissue Equivalent material is independent of the encapsulating material. This is also demonstrated experimentally. © 2001 Elsevier Science Ltd. All rights reserved.
Original languageEnglish
Pages (from-to)415 - 416
Number of pages2
JournalRadiation Physics and Chemistry
Volume61
Issue number3-6
DOIs
Publication statusPublished - 2001
Externally publishedYes