Spatial resolution of the JET Thomson scattering system

L. Frassinetti, M.N.A. Beurskens, R. Scannell, T.H. Osborne, J. Flanagan, M. Kempenaars, M. Maslov, R. Pasqualotto, M. Walsh

Research output: Contribution to journalReview article

52 Citations (Scopus)

Abstract

The instrument function of the high resolution Thomson scattering (HRTS) diagnostic in the Joint European Torus (JET) has been calculated for use in improved pedestal profile analysis. The full width at half maximum (FWHM) of the spatial instrument response is (22 ± 1) mm for the original HRTS system configuration and depends on the particular magnetic topology of the JET plasmas. An improvement to the optical design of the laser input system is presented. The spatial smearing across magnetic flux surfaces is reduced in this design. The new input system has been implemented (from JPN 78742, July 2009) and the HRTS instrument function corresponding to the new configuration has been improved to approximately FWHM = (9.8 ± 0.8) mm. The reconstructed instrument kernels are used in combination with an ad hoc forward deconvolution procedure for pedestal analysis. This procedure produces good results for both the old and new setups, but the reliability of the deconvolved profiles is greatly reduced when the pedestal width is of the same order as, or less than the FWHM of the instrument kernel. © 2012 EURATOM.
Original languageEnglish
Article number013506
Pages (from-to)-
JournalReview of Scientific Instruments
Volume83
Issue number1
DOIs
Publication statusPublished - Jan 2012
Externally publishedYes

    Fingerprint

All Science Journal Classification (ASJC) codes

  • Instrumentation

Cite this

Frassinetti, L., Beurskens, M. N. A., Scannell, R., Osborne, T. H., Flanagan, J., Kempenaars, M., Maslov, M., Pasqualotto, R., & Walsh, M. (2012). Spatial resolution of the JET Thomson scattering system. Review of Scientific Instruments, 83(1), -. [013506]. https://doi.org/10.1063/1.3673467